Lack of agreement between bioimpedance and continuous thermodilution measurement of cardiac output in intensive care unit patients


Authors: Ben N Barry, Abhiram Mallick, Andrew R Bodenham, Michael Vucevic, N Soni, M Haller, C Zollner, J Briegel, H Forst, J Boldt, T Menges, M Wollbruck, H Hammermann, G Hemplemann, T Mihaljevic, L von Segesser, M Tonz, B Leskosek, B Seifert, R Jenni, M Turina, JM Bland, DG Altman, Y Ginosar, LG Thijs, CL Sprung, R Belardinelli, N Ciampani, C Costantini, A Blandini, A Purcaro, WC Shoemaker, CCJ Wo, MH Bishop, PL Appel, JM van de Water, GR Harrington, X Wang, RS Patil, JD Young, P McQuillan, DE Clarke, TA Raffin, WS Sageman, DE Amundsen, I Ovsyscher, S Furman

Journal: Critical Care (1997)

DOI: 10.1186/cc106

Abstract

Bolus thermodilution is the standard bedside method of cardiac output measurement in the intensive care unit (ICU). The Baxter Vigilance monitor uses a modified thermodilution pulmonary artery catheter with a thermal filament to give a continuous read-out of cardiac output. This has been shown to correlate very well with both the 'gold standard' dye dilution method and the bolus thermodilution method. Bioimpedance cardiography using the Bomed NCCOM 3 offers a noninvasive means of continuous cardiac output measurement and has been shown to correlate with the bolus thermodilution method. We investigated the agreement between the continuous bioimpedance and continuous thermodilution methods, enabling acquisition of a large number of simultaneous measurements. ) appeared fair. The Bomed NCCOM 3 bioimpedance monitor shows poor agreement with the Baxter Vigilance continuous thermodilution monitor in a group of general ICU patients and cannot be recommended for cardiac output monitoring in this situation.

Background:

Bolus thermodilution is the standard bedside method of cardiac output measurement in the intensive care unit (ICU). The Baxter Vigilance monitor uses a modified thermodilution pulmonary artery catheter with a thermal filament to give a continuous read-out of cardiac output. This has been shown to correlate very well with both the 'gold standard' dye dilution method and the bolus thermodilution method. Bioimpedance cardiography using the Bomed NCCOM 3 offers a noninvasive means of continuous cardiac output measurement and has been shown to correlate with the bolus thermodilution method. We investigated the agreement between the continuous bioimpedance and continuous thermodilution methods, enabling acquisition of a large number of simultaneous measurements.

Results:

) appeared fair.

Conclusions:

The Bomed NCCOM 3 bioimpedance monitor shows poor agreement with the Baxter Vigilance continuous thermodilution monitor in a group of general ICU patients and cannot be recommended for cardiac output monitoring in this situation.

Introduction

].

].

Bioimpedance cardiography has been developed over the past 30 years as a noninvasive technique to measure cardiac output. Monitors such as the Bomed Noninvasive Computerized Cardiac Output Monitor (NCCOM 3, Bomed Medical Manufacturing Ltd, Cheshire, UK) are commercially available to measure cardiac output. However, in the United Kingdom and elsewhere they have not achieved widespread usage. The NCCOM 3 uses eight spot electrodes, placed at the root of the neck and chest wall. A constant sinusoidal alternating current (2.5 mA rms, 70 kHz) is passed through the subject's chest and the impedance measured. By measuring the maximum rate of change of thoracic impedance during systole, timed from the electrocardiogram (ECG), the stroke volume is calculated. The Sramek-Bernstein formula is used, which calculates stroke volume as the volume of electrically participating intrathoracic tissue × ventricular ejection time × index of contractility, which is the ratio of the peak rate of change in thoracic bioimpedance and the thoracic fluid index (or total thoracic impedance). Cardiac output is then calculated from the product of heart rate and stroke volume, averaged over 16 cardiac cycles.

We have investigated the correlation between these two methods of continuous cardiac output measurement to determine their suitability for use in critically ill patients in the intensive care unit (ICU).

Methods

We compared the Bomed NCCOM 3 with the Baxter Vigilance in a mixed group of seven ICU patients. All patients required pulmonary artery catheterization on clinical grounds. In two patients, an existing PAC was exchanged for a continuous cardiac output PAC, using the same introducer sheath. An explanation of the use of noninvasive bioimpedence monitoring as part of a research study was given to the patients' relatives and assent was obtained. The primary pathologies of the patients were acute pancreatitis (one), emergency repair of an abdominal aortic aneurysm (two), appendix abscess (one), probable pulmonary embolism (one), cholangitis following cholecystectomy (one) and respiratory failure (one). The study was purely observational, and required no alterations in therapy.

For bioimpedance cardiography, eight standard ECG gel electrodes and, if necessary, two additional electrodes for ECG monitoring were applied, according to directions printed on the NCCOM 3 monitor. Timed data points were saved on a personal laptop computer connected to the NCCOM 3 using the CDDP version SI 4.05 software (CDIc, Irvine, California, USA).

For thermodilution measurements, a continuous cardiac output PAC connected to the Vigilance monitor was inserted via the internal jugular or subclavian vein. Timed data was stored in the patient monitoring system (Hewlett Packard Ltd, Boise, Idaho, USA), using the 'Vue-link' software to connect the two devices. A print-out of cardiac output data at 1-min intervals was obtained at the end of the study period.

Any discrepancy between the clocks on the two monitors was accounted for by noting the times displayed at the start of measurement and allowing for this when pairing the data. In this way we ensured that the paired data points were accurately synchronized.

Body surface area was calculated by each device in order to obtain 'indexed' measurements, by entering the patients' height and weight, estimated if necessary. We verified that both devices produced the same body surface area, so eliminating this source of bias error.

].

Results

O. During the study none of the patients suffered new dysrhythmias requiring treatment or interfering with cardiac output measurement.

). The correlation coefficients for individual patients were -0.25, -0.21, 0.41, 0.25, -0.39, -0.16 and 0.06, respectively.

, respectively. Three of the patients showed a poor precision throughout the measured range of cardiac output. The other four patients showed a fair degree of precision, but with a changing bias, from negative to positive, with increasing cardiac output. There were no clear factors which could be used to predict which group individual patients would fall into.

Discussion

]. Bioimpedance cardiography offers an apparently attractive noninvasive way of estimating cardiac output and obtaining derived haemodynamic parameters. However, this is of no benefit if the information acquired is unreliable, leading to inappropriate management. We investigated the use of bioimpedance in the critically ill to assess whether it can be a reliable method of cardiac output measurement for this group of patients.

The thermodilution method is an indirect measure of cardiac output, but has been shown to correlate well with the gold standard dye dilution method. Moreover, it is the method most commonly used to measure cardiac output in patients in the ICU and upon which much of our understanding of the cardiovascular changes in critical illnesses is based.

]. Correct placement of the eight electrodes is important in obtaining accurate information; in ICU patients this may be hampered by dressing covering internal jugular line sites, thoracotomy wounds and chest drain sites. In this study, the directions for electrode placement detailed on the NCCOM 3 monitor were followed as a closely as practically possible, aiming to reproduce the conditions that would pertain to routine clinical use of the monitor.

]. The presence of the thermal filament in the modified PAC used by the Vigilance monitor may also affect bioimpedance measurements; standard PACs may also affect measurements by the presence of the thermistor wire. In any case, if the bioimpedance data are affected by foreign material in the thorax, this makes the use of the Bomed in ICU patients very problematic.

The design of this study is novel in two ways, giving major advantages over previous studies. Firstly, by comparing two 'continuous' methods of cardiac output measurement, the inevitable errors of synchronization using intermittent methods are virtually eliminated. (Previous studies comparing bioimpedance with thermodilution have needed to average several bolus measurements before or after the acquisition of bioimpedance data.) Secondly, we were able to collect a very large number of simultaneous paired data points from the two methods, averaged over the whole respiratory cycle, enabling a more accurate comparison.

would be subject to an error of up to +54% or -62% at the 95% limits of agreement. Importantly, we were able to assess the changes in measured cardiac output in individual patients and these also showed poor agreement with variable precision and bias. This would indicate that the use of the NCCOM 3 is unlikely to be of value even if a subgroup of patients, in whom the precision is acceptable, could be identified.

The two methods cannot therefore be used interchangeably to monitor cardiac output. Furthermore, therapeutic interventions to improve cardiovascular function that have been shown to be beneficial in patients monitored by the thermodilution technique cannot be similarly applied to patients monitored by bioimpedance cardiography. Bioimpedance cardiography cannot be recommended for use in critically in patients such as this group from a general ICU.

Figures and Tables

= 0.01).

). The degree of bias (measured by the mean) and precision (± 2SD) are shown.

Keywords

  • measurement techniques
  • impedance cardiography
  • thermodilution
  • monitoring
  • cardiac output
Want to explore this topic further?

Ask Mako anything about Lack of agreement between bioimpedance and continuous thermodilution measurement of cardiac output in intensive care unit patients — get instant answers, deeper analysis, and related topics.

Research with Mako

Free with your Surf account

Content sourced from Wikipedia, available under CC BY-SA 4.0.

This content may have been generated or modified by AI. CloudSurf Software LLC is not responsible for the accuracy, completeness, or reliability of AI-generated content. Always verify important information from primary sources.

Report